Focus on your patients, not your monitor. The MONITOR Series patient monitors are the perfect choice for busy departments demanding accurate, reliable performance.
The MONITOR can monitor adult and pediatric patients in virtually every department. The innovative user interface features simplified menus and dedicated function keys for timely setups. The big, bright display with large numerics can be easily seen from across the room.
Criticare's patented end-tidal CO2 technology makes the "8100EP-1" an easy choice for your next anesthesia monitor. Whether you intubate or not, we have accessories that will let you reliably obtain capnography on both adults and pediatric patients.
Number of Leads:3 – or 5 Lead
Lead View:User-selectable; I, II, III, aVR, aVL, aVF, V (5-Lead); I, II, or III (3-Lead)
Gain Selection:x0.5, x1, x2, and x4
Frequency Response:Diagnostic: 0.05 to 100 Hz (-3dB);
Monitor: 0.5 to 40 Hz (-3dB);
ST:0.05 to 40 Hz (-3dB
Electrosurgery Protection:Yes
Defibrillator Protection:Yes
Pacer Detection/Rejection:Yes
ST Accuracy: ±0.2 mm or ±5% of reading, whichever is greater, over heart range of 30 to 120 bpm
ST Range: -6.0 to 6.0 mm
NIBP
Technique:Oscillometric measure upon inflation
Inflation Range:Adult/Pediatric: 0 to 300 mmHg: Neonate: 0 to 150 mmHg
Measurement Cycle:< 40 sec. typical; standard adult cuff
Auto. Measurement:2, 3, 5, 10, 15, 30 min; 1, 2, 4 hour intervals
STAT Mode: 5 minutes
of consecutive readings
Maximum Cuff Pressure: Adult/Pediatric:
300 mmHg; Neonate: 150 mmHg
Resolution:1 mmHG
Transducer Accuracy:±2 mmHg or ±2% of reading, whichever is greater, over full range
Clinical Accuracy:Complies with AAMI
PULSE OXIMETRY
Range: 1% to 99%
Resolution:1%
Accuracy:70% to 99% range: ±2%; 50% to 69% range: ±3%; < 50%: unspecified
Method:Dual wavelength LED